CalCOFI workstations rely on a local area network and NAS data server (//seaserv2) to interact and exchange data at-sea. Essential workstations need to be setup & running on the network before the first CTD cast.

The essential workstations are:

  1. Seasoft (from seabird.com) is used to acquire real-time CTD data.
  2. CELogGPS runs in the background (or may be run on another nearby/KVM PC) so the CTD operator can generate the CTDatDepth event, recording 'GPS lat lon date time event# & activity' when the first bottle is tripped.
  3. After tripping the last, surface bottle & the CTD lands on deck, the deck pressure is written on the console ops form; Seasave data acquistion is stopped; the deck unit is turned off; then CTDBackup is run by pressing <Ctrl><Alt>B (shortcut key mapped in Windows to executes the CTDBackup program). This copies all cast files to a flash drive and generates a base file for the seawater sample log, CESL program, running on the wet-lab tablet computer. Note: any corrections to the cast's .mrk file should be done BEFORE pressing <CTRL><Alt>B ie duplicate markers erased or extra bottles without markers added.
  4. ZTree is used to print the .mrk & .prn files then move the CTD cast files to the archive directory when archiving; printouts are collated & put in the CTD notebook.
  5. Network color printer connectivity required.

Other workstations on the network: