CTD Processing Summary CalCOFI 1407NH CTD Final Data | |||
Download 1407NH CTD raw cast files zipped | Download 1407NH FinalQC CTD + bottle data | ||
General CTD Notes - data acquisition notes, logistics, processing - see below | |||
CTD sensor corrections derived by comparing 4 secs of CTD sensor data (prior to bottle closure) to bottle samples | |||
Dual T, S, & O2 | Primary Sensor | Secondary Sensor | |
Salinity offset (bottle - CTD salinity; > 350m only; Seabird SBE4) | -0.0017 | -0.0004 | |
Oxygen ml/L (dual Seabird SBE43) | y = 1.0303x - 0.0136 R² = 0.9997 |
y = 1.0317x - 0.019 R² = 0.9993 |
|
Oxygen umoles/Kg (dual Seabird SBE43) | y = 1.0316x - 0.465 R² = 0.9997 |
y = 1.0331x - 0.7038 R² = 0.9993 |
|
Single sensors | Linear | Polynomial | |
Nitrate - ISUS 4sec ave voltage vs Bottle Nitrate (Satlantic MBARI-ISUS v2) | y = 27.601x - 4.1081 R² = 0.9966 |
||
Fluorometer - linear & polynomial regressions |
y = 9.7003x - 0.4423 R² = 0.7119 |
y = 16.11x2+1.9576x-0.0151 R² = 0.8334 |
|
![]() |
![]() |
||
![]() |
![]() |
||
![]() |
![]() |
||
General notes: These are cast & final CTD Processing Notes from 1407NH cruise | |||
![]() |
|||
CalCOFI 1407NH General Cast Notes: This cruise occupied 75 stations: 66 standard, & 9 SCCOOS. The station schedule was modified due to naval operations on lines 83 & 80 plus bad weather forecast. Refer to the map for the station order and cruise track. Coastal stations on line 83 & 80 (51, 50.5) & Santa Barbara Basin were done after lines 80 & 77 were completed. Cast 01 - trigger #1 replaced pre-cast since it deck-tested poorly; bottle #9 did not close so the CTD was sent back to 13m & bottle 11 closed Cast 03 - bottle 9 did not trip again; trigger changed post-cast. Cast 04 - winch had to go back down on the upcast after the 100m bottle closure due to a bad wire wrap Cast 07 - bottle 20 didn't close so the CTD went back down and bottle 22 was closed at 7m. Cast 10 - rolling seas Cast 11 - extra bottle tripped at 130m wrong depth, went back down for 140m; bottle #3 did not trip - trigger cleaned post-cast Cast 20 & 21 - sta 90.90 prodo cast & separate 3500m cast with 6 bottles tripped at 3500, two at 650m, 4 at surface - bottle #24 closed upon retrieval at the waterline. Cast 21 CalCOFI Line & Sta number changed from 90.0 90.0 to 90.0 90.1 to distinguish the two casts & bottle samples. Cast 23 - new Temperature sensor #2533 installed on secondary channel. Turns out the original T sensor was fine, the wrong config file (from 1404) was loaded. The result is separate data files for downcast 1407023 & upcast 1407023u. Cast 24 - original secondary temperature sensor reinstalled #5102. Cast 26 - fluorometer spikes at depths 515m, 360m, & 210m Cast 27 - full moon; bottle #5 did not close. Spiney lobster larvae in CUFES (sta 90.37) Cast 28 - shallow 300m cast; altimeter detected bottom at ~55m; calm sunny morning Cast 29 - sta mislabeled 90.35 in hdr, should be 90.30; hdr & hex files corrected using text & hex editors. Cast 33 - sta mislabeled 86.7 in hdr, should be 86.8; hdr & hex files corrected using text & hex editors. Cast 36 - Santa Barbara Basin, calm sunrise; 100-400m stratified layers looked screwy on downcast but verified on upcast; pump status reviewed post-cast and pumps were operational. Nine niskins were tripped between 100 - 400m so we'll review those values as well. Cast 39 - fairly rough and rolling Cast 40 - rough seas, winch speed 30m/min all the way down; extra 20m bottle tripped and ignored. Cast 44 - some rolls but relatively calm again Cast 48 - calm warm night Cast 49 - early morning station, sta 83.90, near Navy Operations area. Cast 50 - sta 80.80, stations out of sequence because of naval operations closing line 80 & 83 inshore (>80). Cast 51 - heading west, on sta 80.90; unusual oxygen layer at ~225m seen on both primary & secondary sensors. Downcast stopped at 495m because smoke was smelt - turned out to be cigar smoke, not the winch. CTD continued to 515m. Cast 52 - primary salinity spike at ~430 on downcast, primary T looks fine so probably primary C; secondary salinity is fine Cast 53 - calm with some occasional rolls Cast 55 - rolly on a flat calm day; ship in trough Cast 56 - wrong cast type selected initially; CTD came up to 230m then back to 270m then to 200m Cast 60 - restarted 2x, first because of primary & secondary sensors disagreement, xmlcon verified as correct; second because as the CTD went down, the secondary sensor values did not change. CTD was recovered, the plumbing back flushed to dislodge any biofouling - nothing obvious but when redeployed, everything looked fine. Cast 64 - mistrip bottle #2, extra bottle tripped at 270 - not sampled Cast 65 - deck tested, trigger #2 cleaned pre-cast; calm night with light rain Cast 66 - calm beautiful sunrise; chl max at 16m but no bottle tripped there Cast 67- wrong xmlcon file used again, 1404 from shallow station loaded. Replaced 1407067.xmlcon with 1407066.xmlcon copied & renamed 1407067.xmlcon Cast 69 - bottle #2 did not close Cast 70 - Santa Barbara Basin station Cast 72 - bright sunny SCCOOS station Cast 75- Pt Dume SCCOOS, last station File notes: Cast 29 - sta mislabeled 90.35 in hdr, should be 90.30; hdr & hex files corrected using text & hex editors. 13Feb2015 |